VHP pass box

VHP PASS BOX

   •    Is designed to perform low temperature surface bio decontamination at room temperature with 35% H2O2 Concentration.

Application: 

  • To enable safe transfer of heat sensitive material into sterile production cleanroom (as they are not autoclavable).   
  • Method of H2O2 Injection - Wet Type  
  • Sterilization Assurance Level: 6 Log Reduction is ensured by our equipment, on all exposed surfaces.

Following operations are involved during Decontamination Cycle             

  • Leak Test Phase   
  • Conditioning (i.e., H2O2 Injection) Phase
  • De-contamination Phase   
  • Aeration / Ventilation Phase.

Disinfection level and the process time required to reach the required disinfection

      Level depends on the following properties of the pass box…

  •  Porosity of the surface materials.   
  •  Shape of the pass box.    
  •  Volume of the pass box. 
  •  Distribution of the aerosol in the pass box. 
  •  Can be supplied along with AHU, De-humidifier and BIBO Unit with exhaust HEPA.

Utility Requirements:
            a.) Power Supply - 1 Phase, 230 V ± 10 V, 50 ± 5 Hz Or - 3 Phase, 415 V ± 10 V, 50 ± 5 Hz.

            b.) Compressed Air     - 0.5 Bar to 4 Bar.

  • Overall dimension of our equipment is customized, based on the site / client requirement.

Key Features:  

  • Design is in line with CGMP and GAMP5 requirements.
  • In compliance with 21 CFR Part 11 requirements.  
  • Equipped with necessary Safety and Process Interlocks.    
  • Multiple recipes / cycles can be created.  
  • Exclusive Operating (IPC) panel and Monitoring (HMI) panel are supplied (if need be).
  • Easily operable.

 

 

 

 